Features:
*Dual Technology Features Automatic Selection of Magnetic Induction or Eddy Current Measurement Techniques
*Non-magnetic coatings(e.g. paint, zinc)on steel *Insulating coatings(e.g. paint)on no-ferrous metals
*No-ferrous metals coatings on insulating substrates
*Menu operation system
*Two measuring mode: CONTINUE and SINGLE mode
*Two working Mode: DIRECT and GROUP mode
*Statistic Display: AVG, MAX, MIN, NO., S.DEV
*One point calibrating, two points calibrating and basic calibrating easily
*Memory for 400 readings(80 DIRECT and 320 GROUP readings)
*Delete single readings and all group readings easily
*High alarm and Low alarm for all working mode
*Low battery, error indication
*USB interface for PC software(windows 98/2000/xp/vista)
Specifications:
Sensor probe |
F |
N |
Working principle |
magnetic induction |
Eddy current principle |
Measuring range |
0~1250um |
0~1250um |
Guaranteed tolerance (of reading) |
(+/- 3%+1)um |
(+/- 3%+1.5)um |
Low range Precision |
0.1um |
0.1um |
Minimum curvature radius |
1.5mm |
3mm |
Diameter of Minimum area |
3mm |
5mm |
Basic critical thickness |
0.5mm |
0.3mm |
Size(HxWxD) |
113.5mm x 54mm x 27mm |
Weight |
110g |
Standard configurations:
Main unit
Two ¡°AAA¡± batteries,
USB cable,
Software CD,
Calibration Iron,
Calibration Aluminum,
Precision Standard,
carrying case. |