Search£º
Categories
 
Information
 
Coating Thickness Gauges
CTG2600 Coating Thickness Gauge
Features:

*Two measuring methods: magnetic induction (F) and eddycurrent (N)
*2 measurement modes: continuous / single
*5 types of probes can be used (F400, F1, F10, N1, CN02,)£»
*5 statistical ways: Mean values / Max. values / Min. values / testing numbers.(No.) / standard deviations (S.DeV)
*5 statistic values£ºmean value£¨MEAN£©, max. value£¨MAX£©, min. value£¨MIN£©, numbers of measuring£¨NO£®£©, standard deviation£¨S£®DEV£©£»
*Direct testing mode and block statistics mode (APPL/BATCH)
*Direct print out of statistical values
*Low battery indication
*2 switch off modes: manual and auto
*The instrument caters for the demands as follows:GB/T 4956©¤1985 GB/T 4957©¤1985 JB/T 8393©¤1996 JJG 889©¤95 JJG 818©¤93
*Application fields : Electroplating industry, Spray-painting and plastic-painting industry ,Aluminum processing industry, Pipeline industry,
*Car manufacture industry Steel construction and color steel plate industry
*The CTG2600 dataview software with USB interface is in the standard delivery for the customer
 
Specifications:
Measuring range
Refer to the below table1¡¢table 2
Probes available
Tolerance
Minimum resolution:
Measuring condition
Operation language
English
Standards
DIN, ISO, ASTM,BS
Calibration
Zero and foil calibration
Statistics
Number of measurements, mean, standard deviation,maximum and minimum of 3000 readings
Data memory
495 readings
Humidity:
20%¡«90%RH
Environment of application:
0-40¡æ (No strong magnetic field)
power supply
2¡Á1.5V AA
Dimensions
1250mm¡Á67 mm¡Á31 mm
Weight
400g
 
Standard configurations:
Main unit
1
Probe
1
Standard test plate
5
Substrate
1
Carrying Case
1
Operation Manual
1

Optional probes and technical specification

Table 1
Probe model
F400
F1
F1/90¡ã
F5
F10
Operating principle
Magnetic induction
Measuring range (mm)
0¡«400
0¡«1250
0¡«5000
0¡«10000
Low range resolution(mm)
0.1
0.1
1
10
Accuracy
One-point calibration (mm)
¡À(3%H+1)
¡À(3%H+5)
¡À(3%H+10)
Two-point calibration (mm)
¡À((1¡«3)%H+0.7)
¡À((1¡«3)%H+1)
¡À((1¡«3)%H+5)
¡À((1¡«3)%H+10)
Measuring conditions
Min curvature radius (mm)
Convex
1
1.5
Flatten
5
10
Diameter of the min area (mm)
F3
F7
F7
F20
F40
Critical thickness of substrate (mm)
0.2
0.5
0.5
1
2
 Table 2
Probe model
N400
N1
N1/90¡ã
CN02
N10
Operating principle
Eddy current
Measuring range (mm)
0¡«400
(Chrome on copper
0¡«1250
0¡«40)
10¡«200
0¡«10000
Low range resolution(mm)
0.1
0.1
1
10
Accuracy
One-point calibration (mm)
¡À(3%H+0.7)
¡À(3%H+1.5)
¡À(3%H+1)
¡À(3%H+25)
Two-point calibration (mm)
¡À[(1¡«3)%H+0.7]
¡À[(1¡«3)%H+1.5]
___
¡À[(1¡«3)%H+25]
Measuring conditions
Min curvature radius (mm)
Convex
1.5
3
Flatten
Only flatten
25
Diameter of the min area (mm)
F4
F5
F5
F7
F50
Critical thickness of substrate (mm)
0.3
0.3
0.3
No limit
50mm aluminum foil
 
Beijing TIME Haofang Technology Co., Ltd. All Rights Reserved
Tel: +86-10-61199378 Fax: +86-10-61199378 E-mail:1829881916@qq.com ¾©ICP±¸11029437ºÅ-2 Sign in